產品明細
HM8118
LCR電橋
產品簡述
♦ 基本精度: 0.05%
♦
高達每秒12次量測
♦
量測功能: L, C, R, |Z|, X, |Y|, G, B, D, Q, Φ, ∆, M, N
♦
頻率範圍: 20Hz~200kHz
♦ 內部可進行程式設計的電壓與電流偏置(bias)
♦ 包含Kelvin纜線和四端SMD測試轉接器
詢價
產品詳細介紹
Product Introduction
The R&S®HM8118 provides the full range of measurements to characterize resistors, capacitors and inductors; results are displayed in absolute, relative or average values.
Features
Basic Accuracy 0.05 %
Measurement Functions L, C, R, |Z|, X, |Y|, G, B, D, Q, Θ, ∆, M, N
Test Frequencies 20 Hz to 200 kHz
Up to 12 Measurements per Second
Parallel and Serial Mode
Binning Interface HO118 (optional) for automatic Sorting of Components
Internal programmable Voltage and Current Bias
Transformer Parameter Measurement
External Capacitor Bias up to 40 V
Kelvin Cable and 4-Terminal SMD Test Adapter included
Galvanically isolated USB/RS-232 Dual-Interface, optional IEEE-488 (GPIB)
Specifications
200kHz LCR-Bridge R&S®HM8118
All data valid at 23°C after 30 minutes warm-up.
Conditions
Test signal voltage
1V
Open and short corrections performed
Measurement time
SLOW
Display
Measurement modes
Auto, L-Q, L-R, C-D, C-R, R-Q, Z-Θ, Y-Θ, R-X, G-B, N-Θ, M
Equivalent circuits
auto, series or parallel
Parameters displayed
Value, deviation or % deviation
Averaging
2 to 99 measurements
Accuracy
Primary parameters
Basic accuracy (Test voltage: 1.0V, measurement SLOW/MEDIUM, autoranging mode, constant voltage OFF, bias off). For FAST mode double the basic accuracy values
Impedance:
Secondary parameters
Basic accuracy D, Q
±0,0001 if f = 1kHz
Phase angle
±0,005° if f = 1kHz
Ranges
|Z|, R, X
0,01mΩ to 100MΩ
|Y|, G, B
10nS to 1.000S
C
0,01pF to 100Mf
L
10nH to 100kH
D
0,0001 to 9,9999
Q
0,1 to 9.999,9
θ
-180° to +180°
∆
-999,99 to 999,99%
M
1µH to 100H
N
0,95 to 500
Measurement conditions and functions
Test frequency
20Hz to 200kHz (69 steps)
Frequency accuracy
±100ppm
AC test signal level
50mVrms to 1.5Vrms
Resolution
10mVrms
Drive level accuracy
±(5% + 5mV)
Internal bias voltage
0VDC to +5,00VDC
Resolution
10mV
External bias voltage
0VDC to +40VDC (fused 0.5A)
Internal bias current
0mA to +200mA
Resolution
1mA
Range selection
Auto and Hold
Trigger
Continuous, manual or external via interface, binning interface or trigger input
Trigger delay time
0ms to 999ms in 1ms steps
Measurement time
(f ≥1kHz)
FAST
70ms
MEDIUM
125ms
SLOW
0.7s
Miscellaneous
Test signal level monitor
Voltage, current
Error correction
Open, short, load
Save/Recall
9 instrument settings
Front-end protection
Vmax <√2/C @Vmax <200V, C in Farads (1 Joule of stored energy)
Low potential and low current guarding
Ground, driven guard or auto (fused)
Constant voltage mode (25Ω source)
Temperature effects R, L or C
±5ppm/°C
Interface
Dual interface USB/RS-232 (R&S®HO820), optional R&S®HO880 IEEE-488 (GPIB)
Safety
Safety class I (EN61010-1)
Power supply
110V to 230V ±10%, 50 to 60Hz, CAT II
Power consumption
approx. 20W
Operating temperature
+5°C to +40°C
Storage temperature
-20°C to +70°C
Rel. humidity
5% to 80% (non condensing)
Dimensions (W x H x D)
285 x 75 x 365mm
Weight
approx. 4kg