產品明細

HMC8015

功率分析儀

    產品簡述
    ♦ 用於交流/直流負載電流及待機電流的特性量測
    符合IEC62301, EN50564, 以及EN61000-3-2的性能和一制性協定
    ♦ 基本精確度: 0.05%
    頻寬範圍: DC ~ 100kHz
    功率範圍: 50uW ~ 12kW
    取樣率: 500ksample/s

      產品詳細介紹Product Introduction

      The R&S®HMC8015 Power analyzer is a compact tester for AC/DC load and standby current characterization that enables measurements without additional tools such as a computer or remote infrastructure. In addition to a numerical and graphical display with 26 key parameters, the instrument delivers performance and compliance protocols in line with IEC 62301, EN 50564 and EN 61000-3-2.

      Features
      • Power measurement range: 50 µW to 12 kW
      • Analog bandwidth: DC to 100 kHz
      • Sampling rate: 500 ksample/s
      • Basic accuracy: 0.05 %

      Specifications
      Specifications in brief
      Analog bandwidthDC to 100 kHz
      Frequency accuracy0.1% of reading
      A/D converter resolutionvoltage and current16 bit
      Basic accuracy0.05% of reading
      Display resolution5 digits, 10 updates/s
      Sampling frequency 500 kHz
      Input impedance2 MΩ
      PLL synchronization sourcesU, I, external
      Filters
        Analog signal filter1 kHz
        Digital filterdepends on measurement frequencyadaptive filter
        Frequency filter independent of analog signal filter 500 Hz
      Additional inputs/outputsBNC, rear panel
        Analog input ±10 V (U peak)
        Analog input accuracy0.5% of reading
        Analog output±5 V (U peak)
      Digital input
        Limits (PASS/FAIL)bandwidth100 kHz
      PWM1 kHz
      state≤ 10 Hz
        Signal thresholdlogical zero0 V to 2 V
      logical one3 V to 24 V
      Digital outputup to 100 mA source/sink5 V, TTL
      Display8.9 cm (3.5") TFT (color)
      Resolution320 × 240 pixel (QVGA)
      Network connectionAC100 V to 115 V/230 V at 50 Hz to 60 Hz
      Power consumptionmax. 35 W, typ. 15 W
      Operating temperature range+5°C to +40°C
      Storage temperature range–25°C to +60°C
      Standards CAN/CSA-C22.2 NO. 61010-1, UL 61010-1, DIN EN61010-1, DIN EN61326-1, DIN EN55011
      Common mode voltage CAT II, 600 V (U RMS)
      Dimensions222 mm × 88 mm × 280mm
      (8.74 in × 3.46 in × 11.02 in)
      Weightapprox. 3.25 kg (7.16 lb)
      Warm-up time60 min
      All specifications refer to a sine reference signal, PF = 1, voltage to ground = 0 V, analog filter
      deactivated, digital filters activated and are valid for measurement values > 1% of measurement range.