產品明細

NGP800

電源供應器

    產品簡述
    ♦ 最大輸出功率: 800W
    每通道最大輸出電壓: 32V或64V
    每通道輸出最大電流: 10或20A
    電壓最高達250V(串聯), 電流最高達80A(並聯)
    ♦ 觸控大螢幕

      產品詳細介紹Product Introduction

      Boost your efficiency with quadcore power

      The R&S®NGP800 DC power supply series, comprising five models with 400 W or 800 W, provides maximum power at a variety of operating points. The two or four 200 W outputs can each supply up to 64 V or up to 20 A. Electrically equivalent and galvanically isolated outputs can be wired in series or parallel for up to 250 V or 80 A. Synchronizing your outputs, performing waveform tests and logging data for in-depth analysis – all this becomes easy with the R&S®NGP800 power supply series
      An intuitive operating concept and a large touchscreen allow you to enter values much faster and to display statistics in real time. All R&S®NGP800 power supplies include sense terminals, USB and a LAN interface. A user-installable GPIB interface, a digital trigger I/O, an analog input and a wireless LAN interface are optional, making these instruments great on the bench or in an automated test system.


      Available models
      ModelsNumber of output channelsMax. output powerOutput voltage per channelOutput current per channel
      R&S®NGP8022400 W0 V to 32 V0 A to 20 A
      R&S®NGP8044800 W0 V to 32 V0 A to 20 A
      R&S®NGP8144800 WCh 1, Ch 2: 0 V to 32 V
      Ch 3, Ch 4: 0 V to 64 V
      Ch 1, Ch 2: 0 A to 20 A
      Ch 3, Ch 4: 0 A to 10 A
      R&S®NGP8222400 W0 V to 64 V0 A to 10 A
      R&S®NGP8244800 W0 V to 64 V0 A to 10 A
      Features
      • Full flexibility
        • 5" high-resolution touch display
        • FlexPower
        • Four power supplies in a single instrument
        • Parallel and serial operation

      • Full functionality
        • Ramp function
        • Output delay
        • Arbitrary function
        • Remote sensing
        • Built-in measurements
        • Data logging

      • Full safety
        • Protection functions
        • Safety limits
        • Safe working environment

      • Full connectivity
        • Digital remote control
        • Digital trigger I/O
        • Analog input

      Specifications
      Electrical specifications
      OutputsThe channel outputs are galvanically isolated and not connected to ground.
      Number of output channelsR&S®NGP802, R&S®NGP8222
      R&S®NGP804, R&S®NGP824, R&S®NGP8144
      Total output powerR&S®NGP802, R&S®NGP822max. 400 W
      R&S®NGP804, R&S®NGP824, R&S®NGP814max. 800 W
      Maximum output power per channel200 W
      Output voltage per channelR&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2)0 V to 32 V
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4)0 V to 64 V
      Maximum output current per channel R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2)20 A
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4)10 A
      Maximum voltage in serial operationR&S®NGP80264 V
      R&S®NGP822, R&S®NGP804, R&S®NGP814128 V
      R&S®NGP824250 V
      Maximum current in parallel operation R&S®NGP82220 A
      R&S®NGP802, R&S®NGP824, R&S®NGP81440 A
      R&S®NGP804 80 A
      Voltage ripple and noise 20 Hz to 20 MHz < 3 mV (RMS),
      < 30 mV (Vpp) (meas.)
      Current ripple and noise 20 Hz to 20 MHz< 3.5 mA (RMS) (meas.)
      Load regulationload change: 10% to 90%
      Voltage±(% of output + offset)
      R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2)< 0.01% + 5 Mv
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4)< 0.01% + 10 mV
      Current±(% of output + offset)< 0.01% + 5 mA
      Load recovery time 50% to 100% load change to within 0.2% of rated voltage< 400 µs (meas.)
      Rise time10% to 90% of rated output voltage, resistive load
      R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2)< 10 ms
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) < 12 ms
      Fall time 90% to 10% of rated output voltage, resistive load
      R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2)full load: < 10 ms, no load: < 50 ms
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4)full load: < 25 ms, no load: < 50 ms
      Programming resolution
      Voltage1 mV
      Current0.5 mA
      Programming accuracy
      Voltage±(% of setting + offset)
      R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2)< 0.05% + 5 mV
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4)< 0.05% + 10 mV
      Current ±(% of setting + offset)
      < 0.1% + 5 mA
      Output measurements
      Measurement functionsvoltage, current, power, energy
      Readback resolution
      Voltage1 mV
      Current0.5 mA
      Readback accuracy
      Voltage±(% of output + offset)
      R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2)< 0.05% + 5 mV
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4)< 0.05% + 10 mV
      Current±(% of output + offset)
      < 0.1% + 5 mA
      Temperature coefficient (per °C) ±(% of output + offset), +5°C to +20°C and +30°C to +40°C
      R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) voltage: < 0.0075% + 0.75 mV, current: < 0.015% + 0.75 mA
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) voltage: < 0.0075% + 1.5 mV, current: < 0.015% + 0.75 mA
      Remote sensing
      Maximum sense compensation1 V (meas.)
      Ratings
      Maximum voltage to ground250 V DC
      Maximum countervoltage voltage with the same polarity connected to the outputs
      R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2)35 V
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4) 70 V
      Maximum reversevoltage voltage with opposite polarity connected to the outputs0.4 V
      Maximum reverse current for 5 min max.20 A
      Remote control
      Command processing time < 6 ms (typ.)
      Protection functions
      Overvoltage protectionadjustable for each channel
      Programming resolution1 mV
      Overpower protectionadjustable for each channel
      Overcurrent protection (electronic fuse) adjustable for each channel
      Programming resolution0.5 mA
      Response time(Iload > Iresp × 2) at Iload ≥ 2 A < 1 ms
      Fuse linking (FuseLink function) Yes
      Fuse delay at output-onadjustable for each channel10 ms to 10 s (1 ms increments)
      Fuse delay timeadjustable for each channel10 ms to 10 s (1 ms increments)
      Response time for linked channels< 5 ms
      Overtemperature protectionindependent for each channel
      Special functions
      Output ramp function EasyRamp
      EasyRamp time10 ms to 60 s (1 ms increments)
      Output delay
      Synchronicity< 1 ms (typ.)
      Delay per channel10 ms to 10 s (1 ms increments)
      Arbitrary function QuickArb
      Parametersvoltage, current, time
      Maximum number of points1024
      Maximum number of subgroups8
      Dwell time1 ms to 60 s (1 ms increments)
      Repetitioncontinuous or burst mode with 1 to 65535 repetitions
      Triggermanually, by remote control or via optional trigger input
      Trigger and control interfacesR&S®NGP-K103digital I/O, 16-pin connector block
      Trigger response time< 3 ms (typ.)
      Maximum voltage (IN/OUT)5.5 V
      Input trigger levelTTL
      Maximum drain current (OUT)5 mA
      Analog control interface R&S®NGP-K107analog input, 16-pin connector block
      Input voltage0% to 100% control of voltage or current0 V to 5 V
      Output accuracyR&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2) voltage: < 0.1% + 16 mV,
      current: < 0.1% + 30 mA
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4)voltage: < 0.1% + 32 mV,
      current: < 0.1% + 15 mA
      Temperature coefficient(per °C) ±(% of output + offset), +5°C to +20°C and +30°C to +40°C
      R&S®NGP802, R&S®NGP804, R&S®NGP814 (CH1, CH2)voltage: < 0.015% + 2.4 mV,
      current: < 0.015% + 4.5 mA
      R&S®NGP822, R&S®NGP824, R&S®NGP814 (CH3, CH4)voltage: < 0.015% + 4.8 mV,
      current: < 0.015% + 2.25 mA
      Update rate1 ms
      Data logging
      Maximum acquisition rate 125 sample/s
      Memory depth800 Mbyte internal or external memory
      Voltage resolutionsee readback resolution
      Voltage accuracysee readback accuracy
      Current resolutionsee readback resolution
      Current accuracysee readback accuracy
      Display and interfaces
      Display TFT 5" 800 × 480 pixel WVGA touch
      Front panel connections4 mm safety sockets
      (channel outputs, remote sensing)
      Rear panel connectionsR&S®NGP802, R&S®NGP8228-pin connector block
      (channel outputs and remote sensing)
      R&S®NGP804, R&S®NGP824, R&S®NGP8142 × 8-pin connector block
      (channel outputs and remote sensing)
      Remote control interfaces StandardUSB-TMC, USB-CDC (Virtual COM), LAN
      R&S®NGP-K102 WLAN
      R&S®NG-B105IEEE-488 (GPIB)
      General data
      Environmental conditions
      Temperatureoperating temperature range+5°C to +40°C
      storage temperature range–20°C to +70°C
      HumidityNoncondensing5% to 95%
      Power rating
      Mains nominal voltage100 V to 250 V
      Mains frequency50 Hz to 60 Hz
      Maximum power consumptionR&S®NGP802, R&S®NGP822650 W
      R&S®NGP804, R&S®NGP824, R&S®NGP8141125 W
      Mains fusesinternal (not user accessible)16 A 250 V IEC60127-2/7 fast acting
      Product conformity
      Electromagnetic compatibilityEU: in line with Radio Equipment Directive 2014/53/EUapplied standards:
      ► ETSI EN300328 V2.1.1
      ► EN61326-1
      ► EN61326-2-1
      ► EN55011 (Class A)
      ► EN55032 (Class A)
      ► ETSI EN301489-1 V2.2.0
      ► ETSI EN301489-17 V3.2.0
      KoreaKC mark
      USA, CanadaFCC47 CFR Part 15B, ICES-003 Issue 6
      Electrical safetyEU: in line with Low Voltage Directive 2014/35/EUapplied harmonized standards: EN61010-1
      USA, CanadaUL61010-1, CSA C22.2 No. 61010-1
      WLAN approvalsAustria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United KingdomCE
      SingaporeiMDA standards DB102020
      USA, CanadaFCC, IC
      RoHSin line with EU Directive 2011/65/EUEN50581
      Mechanical resistance
      VibrationSinusoidal5 Hz to 55 Hz, 0.3 mm (peak-to-peak), 55 Hz to 150 Hz, 0.5 g const., in line with EN60068-2-6
      Random8 Hz to 500 Hz, acceleration: 1.2 g (RMS), in line with EN60068-2-64
      Shock40 g shock spectrum, in line with MIL-STD-810E, method 516.4, procedure I
      Mechanical data
      DimensionsW × H × D 362 mm × 100 mm × 451 mm
      (14.25 in × 3.94 in × 17.76 in)
      Weight R&S®NGP802, R&S®NGP822 7.5 kg (16.5 lb)
      R&S®NGP804, R&S®NGP824, R&S®NGP8148.0 kg (17.6 lb)
      Rack installation R&S®ZZA-GE2319 in, 2 HU
      Recommended calibration intervaloperation 40 h/week over entire range of specified environmental conditions1 year