產品明細

FPL1000

頻譜分析儀

    產品簡述
    ♦ 頻率範圍: 5kHz~3GHz/7.5GHz
    ♦ 相位雜訊: <108dBc(1Hz)
    ♦ 使用前置放大器時DANL: < -166dBm (typ.)
    ♦ 內建信號產生器 (選配)
    ♦ 40MHz分析頻寬 (選配)
    ♦ 提供電池/直流電源供應器 (選配), 適用於實驗室和外場應用

      產品詳細介紹Product Introduction

      The R&S®FPL1000 signal and spectrum analyzer makes measuring fast and simple. The intuitive touchscreen is straightforward and easy to use. With its solid RF performance, light weight and small footprint, the R&S®FPL1000 combines the functionality of a benchtop instrument with the portability of a handheld instrument.

      In an RF lab, the R&S®FPL1000 is as indispensable as an oscilloscope or multimeter. It is a single measuring instrument for a variety of measurement tasks. It supports not only spectrum analysis, but also highly accurate power measurement with power sensors and analysis of analog and digitally modulated signals.

      The R&S®FPL1000 is the only instrument in its class that features an internal generator (R&S®FPL1-B9 option) and can analyze signals with a bandwidth of 40 MHz (R&S®FPL1-B40 option).

      The solid RF performance makes the R&S®FPL1000 the ideal instrument for use in the lab, production and service. The 1 dB attenuator step size (R&S®FPL1-B25 option) allows you to perform measurements at the instrument’s maximum dynamic range. The preamplifier (R&S®FPL1-B22 option) extends the sensitivity level. Thanks to its high sensitivity and low phase noise performance, even small interfering signals next to the carrier can be analyzed.

      Using the R&S®FPL1000 is as intuitive as using a smartphone. Simple swiping gestures adjust the center frequency or the reference level. Two-finger gestures change the span or the displayed power level, while the 10.1" screen with 1280×800 pixel resolution provides a clear display of the signal. Furthermore, the user can freely arrange the layout of the measurement results on the display. Using the MultiView display mode, even different measurement modes can be combined and all the results can be displayed on one screen.

      The R&S®FPL1000 has a depth of only one hand length. It fits into any workplace and leaves enough space for DUTs and other measurement instruments.

      Its light weight and battery operation capability lets you take it anywhere to perform measurements. The optional battery pack provides three hours of operation. Thanks to its rich set of accessories, the R&S®FPL1000 is suitable for field measurements. For transport, a protective hard cover is available as well as a padded carrying bag that allows the instrument to be operated while in the bag. A shoulder harness simplifies portable operation.


      Features
      • Frequency range: 5 kHz to 7.5 GHz
      • SSB phase noise: –108 dBc (1 Hz) at 10 kHz offset (1 GHz carrier)
      • DANL with preamplifier: –166 dBm from 10 MHz to 2 GHz
      • Lightweight with small footprint
      • Battery pack and 12 V/24 V power supply (option)
      • Use with power sensors (option)
      • 40 MHz analysis bandwidth (option)
      • Analog and digital signal analysis (option)
      • Internal generator (option)
      • Gated sweep, narrowband resolution filters and spectrogram measurements as standard

      Specifications in brief
      Frequency
      Frequency rangeR&S®FPL10035 kHz to 3 GHz
      R&S®FPL10075 kHz to 7.5 GHz
      Aging per year1 × 10–6
      with R&S®FPL1-B4 option1 × 10–7
      Frequency resolution0.01 Hz
      Bandwidth
      Resolution bandwidth (–3 dB)sweep filters100 kHz to 10 MHz
      in 1/2/3/5 sequence
      FFT filters1 Hz to 50 kHz
      in 1/2/3/5 sequence
      I/Q demodulation bandwidth 12.8 MHz
      with R&S®FPL1-B40 option40 MHz
      Displayed average noise level (DANL)
      RF preamplifier off5 MHz ≤ f < 3 GHz–152 dBm (typ.)
      5 GHz ≤ f ≤ 7.5 GHz–143 dBm (typ.)
      RF preamplifier on
      (option R&S®FPL1-B22)
      10 MHz ≤ f < 2 GHz–166 dBm (typ.)
      5 GHz ≤ f < 7 GHz–159 dBm (typ.)
      Intermodulation
      1 dB compression of input mixer+7 dBm (nom.)
      Third-order intercept point (TOI)300 MHz ≤ fin < 3 GHz+20 dBm (typ.)
      3 GHz ≤ fin ≤ 7.5 GHz+18 dBm (typ.)
      Phase noisef = 1 GHz, 10 kHz frequency offset–108 dBc (1 Hz) (typ.)
      Total measurement uncertainty1 MHz ≤ f < 3 GHz0.5 dB
      3 GHz ≤ f < 7.5 GHz0.8 dB